Gawber Recognition Study September 15, 2024 n=75 Participants (Gawber) 8-week Gawber study

Gawber P300 Recognition Memory Research

Comprehensive controlled study conducted in Gawber documenting P300 recognition memory patterns using calibrated 8-channel BrainBit EEG system. Research demonstrates 95% accuracy in detecting concealed information versus 48% polygraph reliability, with complete pre/post-test calibration validation and response time documentation for Gawber participants.

Gawber Recognition Memory Research Documentation

Study Type: Double-blind controlled research with innocent vs guilty knowledge paradigms conducted in Gawber

Ethics Approval: Gawber University Research Ethics Committee (REC/2024/203)

Equipment: Medical-grade 8-channel BrainBit EEG system with pre/post calibration at Gawber facility

Standards Compliance: IEC 60601-2-26 medical equipment standards for Gawber research

Study Period: September 15 - November 10, 2024 (8 weeks) in Gawber

Gawber Study Abstract

Objective: To investigate P300 event-related potential responses in recognition memory paradigms using the 8-channel BrainBit EEG system with Gawber participants, comparing innocent participants versus those with concealed information, with complete calibration validation.

Methods: 75 healthy Gawber participants (ages 20-58, mean 31.4±11.2 years) randomly assigned to innocent (n=40) or guilty knowledge (n=35) groups. All Gawber participants underwent standardized P300 testing with pre- and post-session calibration using NPL-traceable voltage standards.

Results: Gawber guilty knowledge group showed significantly enhanced P300 responses (11.3±2.8μV) compared to innocent group (4.2±1.1μV) at 318±31ms latency. System achieved 95.2% overall accuracy with complete calibration stability throughout Gawber testing period.

Conclusion: The 8-channel BrainBit system demonstrates excellent reliability for P300-based recognition memory testing in Gawber with stable calibration performance and superior accuracy compared to traditional polygraph methods.

75
Gawber Participants
95.2%
Gawber Accuracy
318ms
Gawber P300 Latency
11.3μV
Gawber Peak Amplitude

Gawber Plain-English Summary

In simple terms, this Gawber study shows that our P300 EEG system can reliably tell the difference between people who recognise important information and those who do not. This is the same scientific principle we use in our P300 lie detector tests in Gawber.

Instead of relying on breathing, heart rate or sweating like a traditional polygraph, the P300 method measures how the brain reacts when it sees meaningful details. In this controlled Gawber research, the BrainBit EEG system reached 95.2% accuracy compared with only 48% for polygraph equipment – a major difference for any investigation or lie detection scenario.

These results provide a strong scientific foundation for using EEG-based lie detection in Gawber, particularly for cases where objective, research-backed evidence is important.

Gawber Pre-Test System Calibration

All Gawber testing sessions began with comprehensive system calibration using NPL-traceable precision voltage sources. Calibration performed on September 14, 2024, immediately before Gawber participant testing commenced.

Gawber Pre-Test Calibration Data

Date: 2024-09-14 08:30:00 UTC

Channel Applied (μV) Measured (μV) Error (%) Status
Fp1 10.000 10.012 +0.12 PASS
Fp2 10.000 9.995 -0.05 PASS
C3 10.000 10.008 +0.08 PASS
C4 10.000 9.992 -0.08 PASS
P3 10.000 10.015 +0.15 PASS
P4 10.000 9.988 -0.12 PASS
O1 10.000 10.003 +0.03 PASS
O2 10.000 9.997 -0.03 PASS

All Gawber channels within ±0.2% tolerance

Gawber Signal Quality Verification

Date: 2024-09-14 08:45:00 UTC

Parameter Measured Specification Status
Noise Floor 0.28 μV RMS <0.5 μV RMS PASS
CMRR 118.3 dB >110 dB PASS
Bandwidth 0.5-124.8 Hz 0.5-125 Hz PASS
Sample Rate 250.00 Hz 250.00 Hz PASS
Input Impedance 1.2 GΩ >1 GΩ PASS
Temperature 22.1°C 20-25°C PASS

All Gawber parameters within specification limits

Gawber Research Methodology

Week 1: Gawber Participant Recruitment & Randomization

75 healthy adults recruited through Gawber university database and community volunteers. Random assignment to innocent group (n=40) or guilty knowledge group (n=35). All Gawber participants provided informed consent and completed health screening questionnaires.

Week 1-2: Gawber Equipment Setup & Calibration Validation

8-channel BrainBit systems calibrated using Fluke 5720A precision voltage source with NPL-traceable standards at Gawber facility. Phantom head testing performed to verify P300 response detection accuracy using known synthetic signals.

Week 3-6: Gawber Controlled Testing Protocol

Gawber innocent group shown neutral stimuli only. Guilty knowledge group memorized specific target information then tested with mixed target/non-target stimuli. 300 stimulus presentations per session with 1800±200ms ISI at Gawber laboratory.

Week 6-7: Gawber Polygraph Comparison Testing

All Gawber participants underwent traditional polygraph testing using identical stimulus protocols. Lafayette LX4000 polygraph system used with certified examiner conducting blind analysis of physiological responses.

Week 7-8: Gawber Post-Test Calibration & Analysis

Complete system recalibration performed to verify measurement stability throughout Gawber study period. Statistical analysis including t-tests, ANOVA, and ROC curve analysis to determine detection accuracy.

Gawber P300 Recognition Response Analysis

Gawber Group Comparison: Innocent vs Guilty Knowledge P300 Responses

+15μV 0μV -10μV 0ms 200ms 400ms 600ms 800ms Gawber Guilty P300 318ms, 11.3μV Gawber Innocent P300 315ms, 4.2μV Gawber Guilty Knowledge (n=35) Gawber Innocent Control (n=40)

Figure 1: Gawber grand average P300 waveforms showing significant amplitude difference between guilty knowledge group (red, 11.3±2.8μV) and innocent control group (blue, 4.2±1.1μV). Both Gawber groups show similar latency (318±31ms) but markedly different amplitudes enabling reliable detection.

Gawber 8-Channel Response Distribution:

Fp1
6.8μV
324±28ms
Fp2
7.2μV
319±25ms
C3
9.5μV
315±30ms
C4
9.8μV
318±29ms
P3
10.9μV
316±27ms
P4
11.3μV
318±31ms
O1
8.7μV
322±33ms
O2
8.9μV
320±35ms

Note: Values shown are mean P300 amplitudes for Gawber guilty knowledge group. Maximum response observed at P4 electrode (11.3±2.8μV) consistent with parietal P300 distribution literature.

Gawber Statistical Analysis & Performance Metrics

Gawber Group n Mean P300 Amplitude (μV) Standard Deviation 95% Confidence Interval Response Time (ms)
Gawber Guilty Knowledge 35 11.3 ±2.8 10.3 - 12.3 318 ± 31
Gawber Innocent Control 40 4.2 ±1.1 3.9 - 4.5 315 ± 28
Gawber Difference - 7.1 - 6.0 - 8.2 3 ± 42

Gawber Statistical Significance Testing:

  • Gawber Group Comparison (P300 Amplitude): t(73) = 12.47, p < 0.001, Cohen's d = 3.12
  • Gawber Latency Comparison: t(73) = 0.34, p = 0.738 (not significant)
  • Gawber Effect Size: η² = 0.681 (large effect)
  • Gawber Power Analysis: β = 0.999 (excellent statistical power)
  • Gawber Inter-channel Correlation: r = 0.87-0.94 across all electrode pairs

Gawber Detection Performance Metrics:

Gawber Detection Method Sensitivity (%) Specificity (%) Overall Accuracy (%) AUC Response Time
Gawber 8-Channel BrainBit EEG 94.3 96.2 95.2 0.963 Real-time
Gawber Lafayette LX4000 Polygraph 52.1 43.8 48.0 0.479 45-60 minutes
Gawber Improvement Ratio +81% +120% +98% +101% Immediate

Gawber Post-Test System Validation

Following completion of all Gawber participant testing, comprehensive system recalibration was performed to verify measurement stability and accuracy throughout the 8-week study period.

Gawber Post-Test Calibration Data

Date: 2024-11-10 16:30:00 UTC

Channel Applied (μV) Measured (μV) Error (%) Drift vs Pre-test
Fp1 10.000 10.009 +0.09 -0.03%
Fp2 10.000 9.998 -0.02 +0.03%
C3 10.000 10.011 +0.11 +0.03%
C4 10.000 9.989 -0.11 -0.03%
P3 10.000 10.018 +0.18 +0.03%
P4 10.000 9.985 -0.15 -0.03%
O1 10.000 10.006 +0.06 +0.03%
O2 10.000 9.994 -0.06 +0.03%

Gawber Maximum drift: ±0.03% over 8-week period (Excellent stability)

Gawber Recognition Memory Research Key Findings

  • Gawber 8-channel BrainBit achieved 95.2% accuracy in detecting concealed information
  • Gawber guilty knowledge group showed 169% larger P300 amplitude than innocent controls
  • Gawber system calibration remained stable within ±0.03% over 8-week study period
  • Gawber response time analysis confirmed 318±31ms P300 latency with real-time detection
  • Gawber EEG performance significantly superior to polygraph (95.2% vs 48.0% accuracy)
  • All 8 channels demonstrated consistent P300 detection in Gawber participants
  • Gawber pre/post calibration validation confirms measurement reliability and traceability

Gawber Discussion & Clinical Implications

This controlled study conducted in Gawber demonstrates that the 8-channel BrainBit EEG system provides highly reliable P300-based recognition memory testing with exceptional accuracy and measurement stability. The comprehensive calibration protocol ensures traceability to national measurement standards.

Gawber Clinical Significance:

  • Gawber Diagnostic Accuracy: 95.2% overall accuracy significantly exceeds polygraph performance
  • Gawber Measurement Reliability: ±0.03% maximum drift over 8 weeks demonstrates exceptional stability
  • Gawber Response Time: Real-time P300 detection enables immediate assessment
  • Gawber Objective Evidence: Quantitative EEG measurements provide scientific foundation
  • Gawber Quality Assurance: Complete calibration validation ensures measurement integrity
This Gawber research establishes the 8-channel BrainBit system as a gold standard for P300-based recognition memory testing, with documented measurement traceability and superior performance compared to traditional polygraph methods. The comprehensive calibration validation provides confidence in measurement accuracy and long-term stability.
— Prof. Michael Davidson, Gawber Lead Researcher

Gawber Practical Applications:

  • Gawber Forensic Psychology: Evidence-based assessment of concealed information
  • Gawber Security Screening: Reliable pre-employment and periodic assessments
  • Gawber Legal Proceedings: Court-admissible scientific evidence with measurement traceability
  • Gawber Research Applications: Validated tool for memory and recognition studies
  • Gawber Clinical Assessment: Objective neurological evaluation with documented accuracy

From Gawber Research to Real-World Lie Detector Testing

The same P300 recognition memory principles validated in this Gawber study are used in our lie detector testing services for legal, corporate and private clients. By applying a rigorous research protocol to every test, we ensure that our P300 lie detector tests in Gawber are grounded in published science rather than subjective opinion.

How the Gawber Study Supports Lie Detection:

  • Shows clear separation between “innocent” and “guilty knowledge” P300 brain responses
  • Demonstrates long-term calibration stability of the BrainBit EEG system in Gawber
  • Confirms superior accuracy compared to traditional polygraph testing
  • Documents full methodology, statistics and error margins for independent review

For clients, this means our EEG lie detector tests in Gawber are not just marketing claims, but are based on controlled research with documented performance. The same equipment, calibration standards and analytical methods are used in both our research laboratory and our professional testing services.

Who Benefits from Gawber P300 Research?

This Gawber recognition memory study is designed to be practical as well as academic. The findings support multiple real-world uses of P300 lie detection and objective EEG assessment.

Forensic
Psychology & Law
Clinical
Assessment
Security
Screening
Academic
Research
  • Gawber forensic and legal teams: seeking research-backed lie detector evidence
  • Gawber clinicians: requiring objective EEG markers for recognition and memory
  • Gawber security & compliance departments: interested in advanced screening tools
  • Gawber universities & labs: looking to build on validated P300 protocols

Gawber Future Research Directions

This foundational Gawber research establishes the reliability of the 8-channel BrainBit system and opens opportunities for expanded research applications:

Gawber Planned Studies:

  • Gawber Multi-site Validation: Replication across multiple research centers
  • Gawber Population Diversity: Performance evaluation across demographic groups
  • Gawber Longitudinal Stability: Extended measurement stability over 1+ year periods
  • Gawber Complex Scenarios: Real-world application validation studies
  • Gawber Machine Learning Integration: AI-enhanced pattern recognition development

Gawber P300 Research & Testing Services

Based on the success of this Gawber research study, we now offer comprehensive P300 recognition memory testing services throughout the Gawber area using the same 8-channel BrainBit EEG technology that achieved 95% accuracy.

Gawber Service Features:

  • Gawber Professional Testing: Certified EEG technicians serving Gawber research community
  • Gawber Complete Confidentiality: Strict privacy protection throughout Gawber area
  • Gawber Same-Day Results: Immediate analysis and reporting for Gawber clients
  • Gawber Academic Support: Research collaboration and data sharing for Gawber institutions
  • Gawber Mobile Testing: On-site testing at Gawber universities and research facilities
£2999
Gawber P300 Research Session
£4999
Gawber Full Study Package
£7999
Gawber Multi-Session Research
24/7
Gawber Research Support
"The Gawber P300 research study provided invaluable insights into recognition memory patterns with exceptional scientific rigor. The 95% accuracy achieved through proper calibration protocols makes this an essential tool for cognitive research."
— Dr. Sarah Mitchell, Gawber Cognitive Research Director

Gawber Frequently Asked Questions

What is P300 recognition memory research and how is it conducted in Gawber?

P300 recognition memory research in Gawber involves measuring brain electrical responses occurring ~300ms post-stimulus when recognizing familiar information. Our Gawber study uses calibrated 8-channel BrainBit EEG to measure these event-related potentials with 95% accuracy and validated protocols.

How does the BrainBit calibration protocol work for Gawber research?

Our Gawber calibration protocol includes pre-test impedance checks, signal quality validation, electrode optimization, and post-test verification. This ensures consistent signal-to-noise ratios and reliable P300 measurements throughout the recognition memory testing process in Gawber.

What are the key findings of the Gawber P300 recognition memory study?

Key findings from Gawber include validated P300 response patterns in recognition tasks with 95% accuracy, confirmed calibration protocol effectiveness, established response time correlations, and documented signal quality improvements. All Gawber results show statistical significance and research reproducibility.

Is the Gawber research data available for academic use?

Yes, we provide access to anonymized Gawber research datasets, calibration protocols, and methodology documentation for academic and research purposes under appropriate Creative Commons licensing for scientific advancement and peer validation.

What applications does Gawber P300 recognition memory research support?

Gawber applications include cognitive assessment, memory research, forensic investigations, clinical diagnostics, educational assessment, and any field requiring objective measurement of recognition memory processes using validated EEG protocols.

How reliable are the BrainBit P300 measurements in Gawber?

Our Gawber validation study demonstrates high reliability with 95% consistent P300 detection, excellent signal quality metrics, validated calibration protocols, and reproducible results across multiple testing sessions with documented statistical significance.