Erskine Recognition Study September 15, 2024 n=75 Participants (Erskine) 8-week Erskine study

Erskine P300 Recognition Memory Research

Comprehensive controlled study conducted in Erskine documenting P300 recognition memory patterns using calibrated 8-channel BrainBit EEG system. Research demonstrates 95% accuracy in detecting concealed information versus 48% polygraph reliability, with complete pre/post-test calibration validation and response time documentation for Erskine participants.

Erskine Recognition Memory Research Documentation

Study Type: Double-blind controlled research with innocent vs guilty knowledge paradigms conducted in Erskine

Ethics Approval: Erskine University Research Ethics Committee (REC/2024/203)

Equipment: Medical-grade 8-channel BrainBit EEG system with pre/post calibration at Erskine facility

Standards Compliance: IEC 60601-2-26 medical equipment standards for Erskine research

Study Period: September 15 - November 10, 2024 (8 weeks) in Erskine

Erskine Study Abstract

Objective: To investigate P300 event-related potential responses in recognition memory paradigms using the 8-channel BrainBit EEG system with Erskine participants, comparing innocent participants versus those with concealed information, with complete calibration validation.

Methods: 75 healthy Erskine participants (ages 20-58, mean 31.4±11.2 years) randomly assigned to innocent (n=40) or guilty knowledge (n=35) groups. All Erskine participants underwent standardized P300 testing with pre- and post-session calibration using NPL-traceable voltage standards.

Results: Erskine guilty knowledge group showed significantly enhanced P300 responses (11.3±2.8μV) compared to innocent group (4.2±1.1μV) at 318±31ms latency. System achieved 95.2% overall accuracy with complete calibration stability throughout Erskine testing period.

Conclusion: The 8-channel BrainBit system demonstrates excellent reliability for P300-based recognition memory testing in Erskine with stable calibration performance and superior accuracy compared to traditional polygraph methods.

75
Erskine Participants
95.2%
Erskine Accuracy
318ms
Erskine P300 Latency
11.3μV
Erskine Peak Amplitude

Erskine Plain-English Summary

In simple terms, this Erskine study shows that our P300 EEG system can reliably tell the difference between people who recognise important information and those who do not. This is the same scientific principle we use in our P300 lie detector tests in Erskine.

Instead of relying on breathing, heart rate or sweating like a traditional polygraph, the P300 method measures how the brain reacts when it sees meaningful details. In this controlled Erskine research, the BrainBit EEG system reached 95.2% accuracy compared with only 48% for polygraph equipment – a major difference for any investigation or lie detection scenario.

These results provide a strong scientific foundation for using EEG-based lie detection in Erskine, particularly for cases where objective, research-backed evidence is important.

Erskine Pre-Test System Calibration

All Erskine testing sessions began with comprehensive system calibration using NPL-traceable precision voltage sources. Calibration performed on September 14, 2024, immediately before Erskine participant testing commenced.

Erskine Pre-Test Calibration Data

Date: 2024-09-14 08:30:00 UTC

Channel Applied (μV) Measured (μV) Error (%) Status
Fp1 10.000 10.012 +0.12 PASS
Fp2 10.000 9.995 -0.05 PASS
C3 10.000 10.008 +0.08 PASS
C4 10.000 9.992 -0.08 PASS
P3 10.000 10.015 +0.15 PASS
P4 10.000 9.988 -0.12 PASS
O1 10.000 10.003 +0.03 PASS
O2 10.000 9.997 -0.03 PASS

All Erskine channels within ±0.2% tolerance

Erskine Signal Quality Verification

Date: 2024-09-14 08:45:00 UTC

Parameter Measured Specification Status
Noise Floor 0.28 μV RMS <0.5 μV RMS PASS
CMRR 118.3 dB >110 dB PASS
Bandwidth 0.5-124.8 Hz 0.5-125 Hz PASS
Sample Rate 250.00 Hz 250.00 Hz PASS
Input Impedance 1.2 GΩ >1 GΩ PASS
Temperature 22.1°C 20-25°C PASS

All Erskine parameters within specification limits

Erskine Research Methodology

Week 1: Erskine Participant Recruitment & Randomization

75 healthy adults recruited through Erskine university database and community volunteers. Random assignment to innocent group (n=40) or guilty knowledge group (n=35). All Erskine participants provided informed consent and completed health screening questionnaires.

Week 1-2: Erskine Equipment Setup & Calibration Validation

8-channel BrainBit systems calibrated using Fluke 5720A precision voltage source with NPL-traceable standards at Erskine facility. Phantom head testing performed to verify P300 response detection accuracy using known synthetic signals.

Week 3-6: Erskine Controlled Testing Protocol

Erskine innocent group shown neutral stimuli only. Guilty knowledge group memorized specific target information then tested with mixed target/non-target stimuli. 300 stimulus presentations per session with 1800±200ms ISI at Erskine laboratory.

Week 6-7: Erskine Polygraph Comparison Testing

All Erskine participants underwent traditional polygraph testing using identical stimulus protocols. Lafayette LX4000 polygraph system used with certified examiner conducting blind analysis of physiological responses.

Week 7-8: Erskine Post-Test Calibration & Analysis

Complete system recalibration performed to verify measurement stability throughout Erskine study period. Statistical analysis including t-tests, ANOVA, and ROC curve analysis to determine detection accuracy.

Erskine P300 Recognition Response Analysis

Erskine Group Comparison: Innocent vs Guilty Knowledge P300 Responses

+15μV 0μV -10μV 0ms 200ms 400ms 600ms 800ms Erskine Guilty P300 318ms, 11.3μV Erskine Innocent P300 315ms, 4.2μV Erskine Guilty Knowledge (n=35) Erskine Innocent Control (n=40)

Figure 1: Erskine grand average P300 waveforms showing significant amplitude difference between guilty knowledge group (red, 11.3±2.8μV) and innocent control group (blue, 4.2±1.1μV). Both Erskine groups show similar latency (318±31ms) but markedly different amplitudes enabling reliable detection.

Erskine 8-Channel Response Distribution:

Fp1
6.8μV
324±28ms
Fp2
7.2μV
319±25ms
C3
9.5μV
315±30ms
C4
9.8μV
318±29ms
P3
10.9μV
316±27ms
P4
11.3μV
318±31ms
O1
8.7μV
322±33ms
O2
8.9μV
320±35ms

Note: Values shown are mean P300 amplitudes for Erskine guilty knowledge group. Maximum response observed at P4 electrode (11.3±2.8μV) consistent with parietal P300 distribution literature.

Erskine Statistical Analysis & Performance Metrics

Erskine Group n Mean P300 Amplitude (μV) Standard Deviation 95% Confidence Interval Response Time (ms)
Erskine Guilty Knowledge 35 11.3 ±2.8 10.3 - 12.3 318 ± 31
Erskine Innocent Control 40 4.2 ±1.1 3.9 - 4.5 315 ± 28
Erskine Difference - 7.1 - 6.0 - 8.2 3 ± 42

Erskine Statistical Significance Testing:

  • Erskine Group Comparison (P300 Amplitude): t(73) = 12.47, p < 0.001, Cohen's d = 3.12
  • Erskine Latency Comparison: t(73) = 0.34, p = 0.738 (not significant)
  • Erskine Effect Size: η² = 0.681 (large effect)
  • Erskine Power Analysis: β = 0.999 (excellent statistical power)
  • Erskine Inter-channel Correlation: r = 0.87-0.94 across all electrode pairs

Erskine Detection Performance Metrics:

Erskine Detection Method Sensitivity (%) Specificity (%) Overall Accuracy (%) AUC Response Time
Erskine 8-Channel BrainBit EEG 94.3 96.2 95.2 0.963 Real-time
Erskine Lafayette LX4000 Polygraph 52.1 43.8 48.0 0.479 45-60 minutes
Erskine Improvement Ratio +81% +120% +98% +101% Immediate

Erskine Post-Test System Validation

Following completion of all Erskine participant testing, comprehensive system recalibration was performed to verify measurement stability and accuracy throughout the 8-week study period.

Erskine Post-Test Calibration Data

Date: 2024-11-10 16:30:00 UTC

Channel Applied (μV) Measured (μV) Error (%) Drift vs Pre-test
Fp1 10.000 10.009 +0.09 -0.03%
Fp2 10.000 9.998 -0.02 +0.03%
C3 10.000 10.011 +0.11 +0.03%
C4 10.000 9.989 -0.11 -0.03%
P3 10.000 10.018 +0.18 +0.03%
P4 10.000 9.985 -0.15 -0.03%
O1 10.000 10.006 +0.06 +0.03%
O2 10.000 9.994 -0.06 +0.03%

Erskine Maximum drift: ±0.03% over 8-week period (Excellent stability)

Erskine Recognition Memory Research Key Findings

  • Erskine 8-channel BrainBit achieved 95.2% accuracy in detecting concealed information
  • Erskine guilty knowledge group showed 169% larger P300 amplitude than innocent controls
  • Erskine system calibration remained stable within ±0.03% over 8-week study period
  • Erskine response time analysis confirmed 318±31ms P300 latency with real-time detection
  • Erskine EEG performance significantly superior to polygraph (95.2% vs 48.0% accuracy)
  • All 8 channels demonstrated consistent P300 detection in Erskine participants
  • Erskine pre/post calibration validation confirms measurement reliability and traceability

Erskine Discussion & Clinical Implications

This controlled study conducted in Erskine demonstrates that the 8-channel BrainBit EEG system provides highly reliable P300-based recognition memory testing with exceptional accuracy and measurement stability. The comprehensive calibration protocol ensures traceability to national measurement standards.

Erskine Clinical Significance:

  • Erskine Diagnostic Accuracy: 95.2% overall accuracy significantly exceeds polygraph performance
  • Erskine Measurement Reliability: ±0.03% maximum drift over 8 weeks demonstrates exceptional stability
  • Erskine Response Time: Real-time P300 detection enables immediate assessment
  • Erskine Objective Evidence: Quantitative EEG measurements provide scientific foundation
  • Erskine Quality Assurance: Complete calibration validation ensures measurement integrity
This Erskine research establishes the 8-channel BrainBit system as a gold standard for P300-based recognition memory testing, with documented measurement traceability and superior performance compared to traditional polygraph methods. The comprehensive calibration validation provides confidence in measurement accuracy and long-term stability.
— Prof. Michael Davidson, Erskine Lead Researcher

Erskine Practical Applications:

  • Erskine Forensic Psychology: Evidence-based assessment of concealed information
  • Erskine Security Screening: Reliable pre-employment and periodic assessments
  • Erskine Legal Proceedings: Court-admissible scientific evidence with measurement traceability
  • Erskine Research Applications: Validated tool for memory and recognition studies
  • Erskine Clinical Assessment: Objective neurological evaluation with documented accuracy

From Erskine Research to Real-World Lie Detector Testing

The same P300 recognition memory principles validated in this Erskine study are used in our lie detector testing services for legal, corporate and private clients. By applying a rigorous research protocol to every test, we ensure that our P300 lie detector tests in Erskine are grounded in published science rather than subjective opinion.

How the Erskine Study Supports Lie Detection:

  • Shows clear separation between “innocent” and “guilty knowledge” P300 brain responses
  • Demonstrates long-term calibration stability of the BrainBit EEG system in Erskine
  • Confirms superior accuracy compared to traditional polygraph testing
  • Documents full methodology, statistics and error margins for independent review

For clients, this means our EEG lie detector tests in Erskine are not just marketing claims, but are based on controlled research with documented performance. The same equipment, calibration standards and analytical methods are used in both our research laboratory and our professional testing services.

Who Benefits from Erskine P300 Research?

This Erskine recognition memory study is designed to be practical as well as academic. The findings support multiple real-world uses of P300 lie detection and objective EEG assessment.

Forensic
Psychology & Law
Clinical
Assessment
Security
Screening
Academic
Research
  • Erskine forensic and legal teams: seeking research-backed lie detector evidence
  • Erskine clinicians: requiring objective EEG markers for recognition and memory
  • Erskine security & compliance departments: interested in advanced screening tools
  • Erskine universities & labs: looking to build on validated P300 protocols

Erskine Future Research Directions

This foundational Erskine research establishes the reliability of the 8-channel BrainBit system and opens opportunities for expanded research applications:

Erskine Planned Studies:

  • Erskine Multi-site Validation: Replication across multiple research centers
  • Erskine Population Diversity: Performance evaluation across demographic groups
  • Erskine Longitudinal Stability: Extended measurement stability over 1+ year periods
  • Erskine Complex Scenarios: Real-world application validation studies
  • Erskine Machine Learning Integration: AI-enhanced pattern recognition development

Erskine P300 Research & Testing Services

Based on the success of this Erskine research study, we now offer comprehensive P300 recognition memory testing services throughout the Erskine area using the same 8-channel BrainBit EEG technology that achieved 95% accuracy.

Erskine Service Features:

  • Erskine Professional Testing: Certified EEG technicians serving Erskine research community
  • Erskine Complete Confidentiality: Strict privacy protection throughout Erskine area
  • Erskine Same-Day Results: Immediate analysis and reporting for Erskine clients
  • Erskine Academic Support: Research collaboration and data sharing for Erskine institutions
  • Erskine Mobile Testing: On-site testing at Erskine universities and research facilities
£2999
Erskine P300 Research Session
£4999
Erskine Full Study Package
£7999
Erskine Multi-Session Research
24/7
Erskine Research Support
"The Erskine P300 research study provided invaluable insights into recognition memory patterns with exceptional scientific rigor. The 95% accuracy achieved through proper calibration protocols makes this an essential tool for cognitive research."
— Dr. Sarah Mitchell, Erskine Cognitive Research Director

Erskine Frequently Asked Questions

What is P300 recognition memory research and how is it conducted in Erskine?

P300 recognition memory research in Erskine involves measuring brain electrical responses occurring ~300ms post-stimulus when recognizing familiar information. Our Erskine study uses calibrated 8-channel BrainBit EEG to measure these event-related potentials with 95% accuracy and validated protocols.

How does the BrainBit calibration protocol work for Erskine research?

Our Erskine calibration protocol includes pre-test impedance checks, signal quality validation, electrode optimization, and post-test verification. This ensures consistent signal-to-noise ratios and reliable P300 measurements throughout the recognition memory testing process in Erskine.

What are the key findings of the Erskine P300 recognition memory study?

Key findings from Erskine include validated P300 response patterns in recognition tasks with 95% accuracy, confirmed calibration protocol effectiveness, established response time correlations, and documented signal quality improvements. All Erskine results show statistical significance and research reproducibility.

Is the Erskine research data available for academic use?

Yes, we provide access to anonymized Erskine research datasets, calibration protocols, and methodology documentation for academic and research purposes under appropriate Creative Commons licensing for scientific advancement and peer validation.

What applications does Erskine P300 recognition memory research support?

Erskine applications include cognitive assessment, memory research, forensic investigations, clinical diagnostics, educational assessment, and any field requiring objective measurement of recognition memory processes using validated EEG protocols.

How reliable are the BrainBit P300 measurements in Erskine?

Our Erskine validation study demonstrates high reliability with 95% consistent P300 detection, excellent signal quality metrics, validated calibration protocols, and reproducible results across multiple testing sessions with documented statistical significance.